Evaluation of Scanner Response with Radiochromic EBT3 Film in Radiation Therapy

Authors

  • Dery TB  Radiological and Medical Sciences Research Institute, Ghana Atomic Energy Commission, Accra, Ghana
  • Amoako JK  Radiation Protection Institute, Ghana Atomic Energy Commission, Accra, Ghana
  • Buah-Bassuah PK  Department of Physics, University of Cape Coast, Cape Coast, Ghana

Keywords:

Scanner; Cobalt-60; EBT3 film; Dosimetry; Calibration

Abstract

The study quantified the performance and evaluated a ?atbed scanner, Epson Stylus CX5900 used for scanning the radiochromic EBT3 films for dosimetry with two other widely used commercial scanners. The performance of each scanner was based on constancy and uniformity. Scanners were tested using film irradiated with doses ranging from 0-500 cGy using 1.25 MeV cobalt-60 isotope. Image J software was used for analysing the scanners. The average dose discrepancy (?) in percentage of the delivered dose was 0.65 % and standard deviation (σ) of 0.92 for the Epson Stylus CX5900 Scanner.

References

  1. Dreindl R., Georg D. and Stock M.Z. (2014). “Radiochromic Film Dosimetry: Considerations on Precision and Accuracy for EBT2 and EBT3 Type Films,” Z. Med. Phys. 24 (2), 153–163.
  2. Zhao L. and Das I.J. (2010). “Gafchromic EBT Film Dosimetry in Proton Beams,” Phys. Med. Biol. 55, 291–301.
  3. Butson M.J., Cheung T., Yu P.K.N. (2005). Absorption Spectra Variations of EBT Radiochromic Film from Radiation Exposure. Phys Med Biol., 50(13), N135–N140.
  4. Chiu-Tsao S.C., Ho Y., Shankar R., Wang L., and Harrison L.B. (2005). Energy Dependence of Response of New High Sensitive Radiochromic Films for Megavoltage and Kilovoltage Radiation Energies. Med Phys., 32(11), 3350–3354.
  5. Arjomandy B., Tailor R., Anand A., Sahoo N., Gillin M., Prado K., and Vicic M. (2010). Radiochromic Film Dosimetry. Medical Physics Publishing, 37, 1942-1947.
  6. Borca V.C., Pasquino M., Russo G., Grosso P., Cante D., Sciacero P., Girelli G., La Porta, M.R., and Tofani S. (2013). Dosimetric Characterization and use of GAFCHROMIC EBT3 Film for IMRT Dose Verification. Journal of Applied Clinical Medical Physics, 14(2), 158-171.
  7. Reinhardt S., Hillbrand M., Wilkens J.J., and Assmann W. (2012). Comparison of Gafchromic EBT2 and EBT3 Films for Clinical Photon and Proton Beams. Med Phys., 39(8), 5257–5262.
  8. León Marroquin E.Y., Herrera González J.A., Camacho López M.A., Villarreal Barajas J.E., García-Garduño O.A. (2016). Evaluation of the Uncertainty in an EBT3 Film Dosimetry System Utilizing Net Optical Density. Journal of Applied Clinical Medical Physics, 17(5).
  9. Niroomand-Rad, A., Blackwell, C. R., Coursey, B. M., Gall, K. P., Galvin, J. M., McLaughlin, W. L., . . . Soares, C. G. (1998). Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists in Medicine. Med. Phys, 25(11), 2093–2115.
  10. Vadrucci M., Esposito G., Ronsivalle C., Cherubini R., Marracino F., Montereali R.M., Picardi L., Piccinini M., Pimpinella M., Vincenti M.A., and De Angelis C. (2015). Calibration of GafChromic EBT3 for Absorbed Dose Measurements in 5 MeV Proton Beam and 60Co?-rays, Medical Physics, 42(8).
  11. Menegotti L., Delana A., and Martignana A. (2008). Radiochromic Film Dosimetry with Flatbed Scanners: A Fast and Accurate Method for Dose Calibration and Uniformity Correction with Single Film Exposure. Medical Physics, 35(7), 3078–3085.
  12. Devic S., Tomic N., Soares C.G., and Podgorsak E.B. (2009). Optimizing the Dynamic Range Extension of a Radiochromic Film Dosimetry System. Med Phys., 36(2), 429–437.
  13. Bouchard H., Lacroix F., Beaudoin G., Carrier J.F., and Kawrakow I. (2009). On the Characterization and Uncertainty Analysis of Radiochromic Film Dosimetry. Med Phys., 36(6), 1931–1946.
  14. Renade M.K., Li J.G., Dubose R.S., Kozelka J., Simon W.E., and Dempsey J.F. (2008). A Prototype Quantitative Film Scanner for Radiochromic Film Dosimetry. Med Phys., 35(2), 473–479.
  15. Martisikova M., Ackerman B., and Jäkel O. (2008). Analysis of uncertainties in Gafchromic EBT Film Dosimetry of Photon Beams. Phys Med Biol., 53(24), 7013–7027.
  16. Paelinck L, De Neve W, De Wagter C. (2007). Precautions and Strategies in Using a Commercial Flatbed Scanner for Radiochromic Film Dosimetry. Phys Med Biol., 52(1), 231–242.
  17. Farah N., Francis Z., and Abboud M. (2014). Analysis of the EBT3 Gafchromic Film Irradiated with 6 MV Photons and 6 Mev Electrons Using Reflective Mode Scanners. Phys Med., 30(6), 708-712. doi: 10.1016/j.ejmp.2014.04.010.

Downloads

Published

2018-04-30

Issue

Section

Research Articles

How to Cite

[1]
Dery TB, Amoako JK, Buah-Bassuah PK, " Evaluation of Scanner Response with Radiochromic EBT3 Film in Radiation Therapy, International Journal of Scientific Research in Science and Technology(IJSRST), Online ISSN : 2395-602X, Print ISSN : 2395-6011, Volume 4, Issue 5, pp.1646-1649, March-April-2018.