Distance between Slices Investigation on Siemens CT Phantom Images

Authors

  • Ida Ayu Sakinah  Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Semarang, Central Java, Indonesia
  • Choirul Anam  Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Semarang, Central Java, Indonesia
  • Eko Hidayanto  Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Semarang, Central Java, Indonesia
  • Ariij Naufal  Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Semarang, Central Java, Indonesia

DOI:

https://doi.org/10.32628/IJSRST523103166

Keywords:

CT SCAN, Siemens PHANTOM, SLICE THICKNESS, FOV, Distance Between Slices

Abstract

This study aims to develop an algorithm for automatic measurement of distance between slices on a Siemens CT phantom with variations of slice thickness and field of view (FOV). The Siemens 64-slice Somatom Perspective CT scanner was utilized to acquire the Siemens phantom image data, with slice thickness variation (i.e., 1, 2, 3, 4, and 5 mm) and FOV variation (i.e., 220, 240, and 300 mm). The distance between slices was calculated by comparing two images from different slice positions. The first stage was opening the images. The second stage was segmentation of the ramp object from image slice-1. The third stage was determination of the centroid coordinates of the ramp object from image slice-1. The fourth stage repeated the processes at second and third stages for slice-2. The fifth stage was calculating the distance between slices. The measured distances were compared with the set distances extracting from DICOM header for every slice position. The developed algorithm can measure distance between slices for variations of slice thickness and FOV. The results of automatic measurement of distance between slices have strong linear correlation ( > 0.99) with the set distances. The differences between automated measurement results and set distances are more than 2 mm for distances up to 50 mm or the differences are around 23% for both slice thickness and FOV variations.

References

  1. Wu RY, Williamson TD, Sahoo N, et al. Evaluation of the high definition field of view option of a large-bore computed tomography scanner for radiation therapy simulation. Phys Imaging Radiat Oncol. 2020;13:44-49.
  2. Lasiyah N, Anam C, Hidayanto E, Dougherty G. Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso-center, and reconstruction filter. J Appl Clin Med Phys. 2021;22(7):313-321.
  3. Roa AM, Andersen HK, Martinsen AC. CT image quality over time: comparison of image quality for six different CT scanners over a six-year period. J Appl Clin Med Phys. 2015;16(2):350-365.
  4. Gulliksrud K, Stokke C, Martinsen AC. How to measure CT image quality: variations in CT-numbers, uniformity and low contrast resolution for a CT quality assurance phantom. Phys Med. 2014;30(4):521-526.
  5. Elnour H, Hassan HA, Mustafa A, et al. Assessment of Image Quality Parameters for Computed Tomoghraphy in Sudan. Scientific Research Publishing. 2017;7(1): 75-84
  6. Greene TC, Rong XJ. Evaluation of techniques for slice sensitivity profile measurement and analysis. J Appl Clin Med Phys. 2014;15(2):281-294.
  7. Samei E & Pfeiffer DE. Clinical Imaging Physics: Current and Emerging Practice. USA: Wiley Blackwell. 2020.
  8. Chadwick JW, Lam EW. The effects of slice thickness and interslice interval on reconstructed cone beam computed tomographic images. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010;110(4):e37-e42.
  9. Tan Y, Guo P, Mann H, et al. Assessing the effect of CT slice interval on unidimensional, bidimensional and volumetric measurements of solid tumours. Cancer Imaging. 2012;12(3):497-505.
  10. Shafiq-Ul-Hassan M, Zhang GG, Latifi K, et al. Intrinsic dependencies of CT radiomic features on voxel size and number of gray levels. Med Phys. 2017;44(3):1050-1062.

Downloads

Published

2023-06-30

Issue

Section

Research Articles

How to Cite

[1]
Ida Ayu Sakinah, Choirul Anam, Eko Hidayanto, Ariij Naufal "Distance between Slices Investigation on Siemens CT Phantom Images" International Journal of Scientific Research in Science and Technology(IJSRST), Online ISSN : 2395-602X, Print ISSN : 2395-6011,Volume 10, Issue 3, pp.931-940, May-June-2023. Available at doi : https://doi.org/10.32628/IJSRST523103166