Simple Automated Verification of Field Size Indicator for Quality Assurance of Medical Linear Accelerator

Authors

  • Siti Hanan  Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Jl. Prof. Soedarto SH, Tembalang, Semarang 50275, Central Java, Indonesia
  • Choirul Anam  Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Jl. Prof. Soedarto SH, Tembalang, Semarang 50275, Central Java, Indonesia
  • Eko Hidayanto  Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Jl. Prof. Soedarto SH, Tembalang, Semarang 50275, Central Java, Indonesia

DOI:

https://doi.org/10.32628/IJSRST22922

Keywords:

Collimator, Field Size Indicator, Linear Accelerator, Quality Assurance, Radiotherapy

Abstract

Purpose: The purpose of this study is to automate the field size verification to facilitate mechanical check aspect medical linear accelerator (linac) quality assurance in a MATLAB-based algorithm on electronic portal imaging device (EPID) images. Methods: A total of 5 reference datasets (i.e. field sizes of 5 cm × 5 cm, 10 cm × 10 cm, 15 cm × 15 cm, 20 cm × 20 cm, and 25 cm × 25 cm) and 15 test datasets (i.e. reference field sizes plus 1 mm, 3 mm, and 5 mm increments) acquired by 6 MV Elekta Linac were used in this study. The proposed algorithm implemented a full automatic threshold with a value of 230 as a segmentation technique. The automated results were compared with manual results obtained using a ruler. Results: The automated results are comparable to manual results (i.e., the difference of both is within 2% or equal to 3 mm). The range of minimum to maximum difference between automated and manual was 0 - 3 mm and the maximum difference found in the 15.3 cm field size setting. Conclusions: We have successfully developed an automated procedure of field size verification and confirmed that the proposed algorithm provide a fast and accurate results.

References

  1. P. Anand et al., “Cancer is a preventable disease that requires major lifestyle changes.” Pharm Res., vol. 25, no. 9, pp. 2097-2116, 2008, doi: 10.1007/s11095-008-9661-9.
  2. T. G. Simon and A. T. Chan, “Lifestyle and Environmental Approaches for the Primary Prevention of Hepatocellular Carcinoma.” Clin Liver Disvol., vol. 24, no. 4, pp. 549-576, 2020, doi: 10.1016/j.cld.2020.06.002.
  3. R. L. Siegel, K. D. Miller, H. E. Fuchs, and A. Jemal, “Cancer Statistics, 2021,” CA Cancer J Clin., vol. 71, no. 1, pp. 7-33, 2021, doi: 10.3322/caac.21654
  4. P. P. Connell and S. Hellman, “Advances in radiotherapy and implications for the next century: a historical perspective,” Cancer Res., vol. 69, no. 2, pp. 383-392, 2009, doi: 10.1158/0008-5472.CAN-07-6871.
  5. S. Gianfaldoni, R. Gianfaldoni, U. Wollina, J. Lotti, G. Tchernev, and T. Lotti, “An Overview on Radiotherapy: From Its History to Its Current Applications in Dermatology,” Open Access Maced J Med Sci., vol. 5, no. 4, pp. 521–525, 2017, doi: 10.3889/oamjms.2017.122
  6. D. I. Thwaites and J. B. Tuohy, “Back to the future: the history and development of the clinical linear accelerator,” Phys Med Biol., vol. 51, no. 13, pp. R343-R362, 2006, doi: 10.1088/0031-9155/51/13/R20.
  7. S. Yani, I. Budiansah, S. Pratama, M. F. Rhani, C. Anam, and F. Haryanto, “Evaluation of the dosimetric characteristics of 10 MV flattened and unflattened photon beams in a heterogeneous phantom,” Int J Radiat Res., vol. 19, no. 4, pp. 835-841, 2021, doi: 10.52547/ijrr.19.4.9.
  8. C. Anam, D. S. Soejoko, F. Haryanto, S. Yani, and G. Dougherty, “Electron contamination for 6 MV photon beams from an Elekta linac: Monte Carlo simulation,” Journal of Physics and Its Applications, vol. 2, no. 2, pp. 97-101, 2020, doi: 10.14710/jpa.v2i2.7771.
  9. G. J. Kutcher et al., “Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40,” Med Phys., vol. 21, no. 4, pp. 581-618, 1994, doi: 10.1118/1.597316.
  10. E. E. Klein et al., “Task Group 142 report: Quality assurance of medical accelerators,” Med. Phys., vol. 36, no. 9, pp. 4197-4212, 2009, doi: 10.1118/1.3190392.
  11. L. Court et al., “Illustrated instructions for mechanical quality assurance of a medical linear accelerator,” J Appl Clin Med Phys., vol. 19, no. 3, pp. 355-359, 2018, doi: 10.1002/acm2.12265.
  12. A. Boyer et al., “AAPM Report No. 72: Basic Application of Multileaf Collimators,” 2001, doi: 10.37206/71.
  13. Y. M. Y. Abdallah and M. A. Boshara, “Assessment of Field Size on Radiotherapy Machines Using Texture Analysis,” Sudan Medical Monitor, vol. 9, no. 1, pp. 5-10, 2014, doi: 10.4103/1858-5000.144642.
  14. C. F. Njeh, B. Caroprese. and P. Desai, “A simple quality assurance test tool for the visual verification of light and radiation field congruent using electronic portal images device and computed radiography,” Radiat Oncol., vol. 7, no. 49, pp. 1-14, 2012, doi: 10.1186/1748-717X-7-49.
  15. D. Létourneau, H. Keller, N. Becker, M. N. Amin, B. Norrlinger, and D. A. Jaffray, “Quality control methods for linear accelerator radiation and mechanical axes alignment,” Med Phys. Vol. 45, no. 6, pp. 2388-2398, 2018, doi: 10.1002/mp.12910.
  16. A. Naufal, C. Anam, C. E. Widodo, and G. Dougherty, “Automated Calculation of Height and Area of Human Body for Estimating Body Weight Using a Matlab-based Kinect Camera,” Smart Science, vol. 10, no. 1, pp. 68-75, 2022, doi: 10.1080/23080477.2021.1983940.
  17. J. W. Wong, “Electronic Portal Imaging Devices (EPID),” In: L. W. Brady, T. E. Yaeger, (eds) Encyclopedia of Radiation Oncology. Springer, Berlin, Heidelberg, 2013, doi: 10.1007/978-3-540-85516-3_33.
  18. B. Sun et al., “Daily QA of linear accelerators using only EPID and OBI,” Med Phys., vol. 42, no. 10, pp. 5584-5594, 2015, doi: 10.1118/1.4929550.

Downloads

Published

2022-04-30

Issue

Section

Research Articles

How to Cite

[1]
Siti Hanan, Choirul Anam, Eko Hidayanto "Simple Automated Verification of Field Size Indicator for Quality Assurance of Medical Linear Accelerator" International Journal of Scientific Research in Science and Technology(IJSRST), Online ISSN : 2395-602X, Print ISSN : 2395-6011,Volume 9, Issue 2, pp.55-60, March-April-2022. Available at doi : https://doi.org/10.32628/IJSRST22922