Evaluation of Tube Current Modulation Profile on CT Image of 128 slices CT Scanner using The Borobudur Phantom
DOI:
https://doi.org/10.32628/IJSRST52310543Keywords:
CT, Siemens, Step wedge, Tube Current ModulationAbstract
The purpose of this study is to evaluate the tube current profile of Siemens 128 slice CT scanner with tube current modulation (TCM) technique using the Borobudur phantom. Borobudur phantom was made from acrylic which has four different diameters of 8, 16, 24, and 32 cm. It was filled with distilled water. The phantom was scanned using a Siemens Somatom 128 slice CT Scanner. The CT scanner took a topogram image first to estimate the attenuation level of each diameter of the phantom. The phantom was scanned using tube voltage and quality reference variations. After the axial images of the phantom were obtained, the tube current information was extracted from digital imaging and communications in medicine (DICOM) header in each dataset using through IndoQCT software. It is found that as the phantom diameter increases, the tube current also increases. For variation of quality reference, the increasing value of quality reference will increase the tube current value. The results show that the tube current profile in the tube voltage variation shows insignificant changes with a statistical test p-value of 0.752. From this study, we conclude that TCM can be evaluated using Borobudur phantom and IndoQCT software.
References
- Paolicchi F, Bastiani L, Negri J, Caramella D. Effect of CT Localizer Radiographs on Radiation Dose Associated With Automatic Tube Current Modulation: A Multivendor Study. Curr Probl Diagn Radiol. 2020;49(1):34-41.
- Brenner DJ. Should we be concerned about the rapid increase in CT usage? Rev Environ Health. 2010;25(1):63-68
- Layman RR, Hardy AJ, Kim HJ, Chou EN, Bostani M, Cagnon C, Cody D, McNitt-Gray M. A comparison of breast and lung doses from chest CT scans using organ-based tube current modulation (OBTCM) vs. Automatic tube current modulation (ATCM). J Appl Clin Med Phys. 2021;22(5):97-109.
- Fazel R, Krumholz HM, Wang Y, Ross JS, Chen J, Ting HH, Shah ND, Nasir K, Einstein AJ, Nallamothu BK. Exposure to low-dose ionizing radiation from medical imaging procedures. N Engl J Med. 2009;361(9):849-857.
- Imai K, Ikeda M, Enchi Y, Niimi T. Quantitative assessment of image noise and streak artifact on CT image: comparison of z-axis automatic tube current modulation technique with fixed tube current technique. Comput Med Imaging Graph. 2009;33(5):353-358.
- Iball GR, Brettle DS, Moore AC. Assessment of tube current modulation in pelvic CT. Br J Radiol. 2006;79(937):62-70.
- Anam C., Haryanto F., Widita R., Arif I., Dougherty G., McLean D. Volume computed tomography dose index (CTDIvol) and size-specific dose estimate (SSDE) for tube current modulation (TCM) in CT scanning. International Journal of Radiation Research. 2018;16(3):289–297.
- Gies M, Kalender WA, Wolf H, Suess C. Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies. Med Phys. 1999;26(11):2235-2247.
- Matsubara K, Sugai M, Toyoda A, Koshida H, Sakuta K, Takata T, Koshida K, Iida H, Matsui O. Assessment of an organ-based tube current modulation in thoracic computed tomography. J Appl Clin Med Phys. 2012;13(2):148-158.
- Kalra MK, Maher MM, Toth TL, Schmidt B, Westerman BL, Morgan HT, Saini S. Techniques and applications of automatic tube current modulation for CT. Radiology. 2004;233(3):649-657.
- Supanich MP., Ford H., Detroit. Strategies for CT-scan parameter optimazation: Tube Current Modulation. Acad. Radiol. 2013:1–7.
- AAPM Alliance for Quality CT. AAPM Computed Tomography Automatic Exposure Control Education Slides. 2021
- Rego SL., Yu L., Bruesewitz MR., Vrieze TJ., et al. CARE Dose4D CT Automatic Exposure Control System : Physics Principles and Practical Hints. Radiol. Soc. North Am. 2007 Sci. Assem. Annu. Meet. 2007.
- Bruesewitz MR., Yu L., Vrieze TJ., Kofler JM., Mccollough CH. Smart mA – Automatic Exposure Control ( AEC ): Physics Principles and Practical Hints Automatic Exposure Control ( AEC ) Systems in CT Using AEC to Prescribe the Desired Image Quality AEC and Image Quality. Med. Educ. 2008.
- AAPM. Performance Evaluation Of Computed Tomography Systems. in AAPM Report No. 223. 2014.
- Merzan D, Nowik P, Poludniowski G, Bujila R. Evaluating the impact of scan settings on automatic tube current modulation in CT using a novel phantom. Br J Radiol. 2017;90(1069):20160308.
- McKenney SE, Seibert JA, Lamba R, Boone JM. Methods for CT automatic exposure control protocol translation between scanner platforms. J Am Coll Radiol. 2014;11(3):285-291.
- Rif’ah SM., Anam C., Sutanto H., Cholid H. Comparison of Water-Equivalent Diameter Measured from CT Localizer Radiograph Based on Two phantoms of the Step-Wedge and Computed Tomography Dose Index. Int. J. Sci. Res. Sci. Technol. 2022;9(6):445–452.
- Anam C, Naufal A, Fujibuchi T, Matsubara K, Dougherty G. Automated development of the contrast-detail curve based on statistical low-contrast detectability in CT images. J Appl Clin Med Phys. 2022 ;23(9):e13719.
- Budi WS. FISIKA SINAR ROENTGEN. Semarang: Undip Press Semarang,2020.
- Samei E, Bakalyar D, Boedeker KL, et al. Performance evaluation of computed tomography systems: Summary of AAPM Task Group 233. Med Phys. 2019;46(11):e735-e756. doi:10.1002/mp.13763
Downloads
Published
Issue
Section
License
Copyright (c) IJSRST

This work is licensed under a Creative Commons Attribution 4.0 International License.