The Feasibility of Tube Current Modulation (TCM) to Reduce dose of the Surface Breast in Various Breast Sizes

Authors

  • Mela Nur Hikmah  Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Jl. Prof. Soedarto SH, Tembalang, Semarang 50275, Central Java, Indonesia
  • Choirul Anam  Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Jl. Prof. Soedarto SH, Tembalang, Semarang 50275, Central Java, Indonesia
  • Heri Sutanto  Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Jl. Prof. Soedarto SH, Tembalang, Semarang 50275, Central Java, Indonesia
  • Oktarina Damayanti  Politeknik al Islam Bandung, Jl. Cisaranten kulon, No.120, Arcamanik, Bandung 40293, West Javat, Indonesia
  • Mohamad Ramdhan  Radiology Department, RSUP dr. Hasan Sadikin, Jl. Pasteur No.38, Pasteur, Sukajadi, Bandung, West Java 40161, Indonesia

DOI:

https://doi.org//10.32628/IJSRST52310682

Keywords:

Computed Tomography, Breast Surface Dose, Noise, Breast Phantom.

Abstract

This study aims to develop three sizes of breast phantoms from silicone rubber (SR) material and evaluate tube current modulation (TCM) to reduce surface doses of the developed breast phantoms. The in-house breast phantom has three sizes: small (cup 34), medium (cup 36), and large (cup 38). The cross-sectional areas of cup 34, 36, and 38 are 78.5, 113.04, and 153.87 cm2, respectively. The in-house phantom was attached to the adult anthropomorphic phantom for dose measurement. Dose measurement was performed using an Optically Stimulated Luminescence (OSL) dosimeter. Scans were performed with and without TCM technique. It was found that the TCM effectively reduce the radiation dose to the breast surface up to approximately 60%. However, the dose reduction was accompanied by a 66% increase in noise when using the TCM technique.

References

  1. Kirby M, Smith BM. Chest. Quantitative CT scan Imaging of the airways for diagnosis and management of lung disease. 2023. Nov; 165(5): 1150-1158. Doi: 10.1016/j.chest.2023.02.004
  2. I.A.E.A. Dose reduction in C.T.C.T. while maintaining diagnostic. 2009 Confidence: a feasibility/demonstration study, IAEA-TECDOC-1621.
  3. Shuryak I, Sachs RK, Brenner DJ. J Natl Cancer Inst.Cancer Risks After Radiation Exposure in Middle Age. 2010. Nov 3; 102(21):1628–1636. Doi: 10.1093/jnci/djq346.
  4. Han MA, Kim JH. (2018) Diagnostic X-Ray Exposure and Thyroid Cancer Risk: Systematic Review and Meta-Analysis. Thyroid.;28(2):220-228.
  5. Mackey SE, Creasman WT. (1995) Ovarian cancer screening. J Clin Oncol.;13(3):783-793. doi:10.1200/JCO.1995.13.3.783
  6. Golubicic I, Borojevic N, Pavlovic T. (2008) Risk factors for breast cancer: is ionizing radiation among them? J BUON.;13(4):487-494.
  7. Ribeiro Guerra M, Coignard J, Eon-Marchais S, et al. Breast cancer. Diagnostic chest X-rays and breast cancer risk among women with a hereditary predisposition to breast cancer unexplained by a BRCA1 or BRCA2 mutation. 2021. August 3;23(1):79. Doi 10.1186/s13058-021-01456-1.
  8. Unnithan J, Macklis RM. Radiother Oncol. Contralateral breast cancer risk. 2001. Sep;60(3):239-46. Doi: 10.1016/s0167-8140(01)00369-3.
  9. Klein L, Liu C, Steidel J, et al. Med Phys. Patient-specific radiation risk-based tube current modulation for diagnostic 2022 Juli;49(7):4391-4403. Doi: 10.1002/mp.15673.
  10. He w, Huda W, Magill D, Tevrides E, Yao H. Radiant Prot Dosimetry. X-ray tube current modulation and patient doses in the chest. 2011 Jan; 143(1):81-7. Doi: 10.1093/rpr/ncp291.Apub 2010 November 9.
  11. Taylor S, Litmanovich DE, Shahrzad M, Bankier AA, Gevenois PA, Tack D. Radiology. Organ-based tube current modulation: are women's breasts positioned in the reduced-dose zone? Radiology.;274(1):260-266. 2015. Jan; 274(1):260-6. Doi: 10.1148/radiol. 14140694. 
  12. Feldle P, Grunz JP, Huflage H, et al. Eur J Radiol. Organ-based tube current modulation versus spectral shaping via tin prefiltration: What does bismuth breast shielding add in low-dose lung C.T.? 2023 Dec:169:111157. Doi: 10.1016/j.ejrad.2023.111157.
  13. Layman RR, Hardy AJ, Kim HJ, et al. Appl Clin Med Phys. A comparison of breast and lung doses from chest C.T scans using organ-based tube current modulation (O.B.T.C.M.) vs. Automatic tube current modulation (A.T.C.M.). 2021 May;22(5):97-109. Doi: 10.1002/acm2.13198. Epub 2021 May 3.
  14. Vollmar SV, Kalender WA. Eur Radiol. Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated C.T.C.T. examinations. Aug;18(8):1674-82. Doi: 10.1007/s00330-008-0934-9.
  15. Tang Y, Tian W, Xie J, et al. Front Immunol. Prognosis and Dissection of Immunosuppressive Microenvironment in Breast Cancer Based on Fatty Acid Metabolism-Related Signature. 2022. Mar 31:13:843515. Doi: 10.3389/fimmu.2022.843515. eCollection
  16. Seidenfuss A, Mayr A, Schmid M, Uder M, Lell MM. A.J.R. Am Roentgenol.Dose reduction of the female breast in chest C.TC.T. May;202(5): W447-52. doi: 10.2214/AJR.13.10541.
  17. Saleh G, Abuelhaija A, Alfaris B, et al. Plos One Heterogeneous breast phantom for computed tomography and magnetic resonance imaging. 2023 Apr 13;18(4): e0284531. Doi: 10.1371/journal.pone.0284531. eCollection 2023.
  18. H.M. Ismail, C.G. Pretty, M.K.M.K. Signal, M. Haggers, C. Zhou, J.G. Chase, Biomed. Phys. Eng. Express 3. Mechanical Behaviour of Tissue Mimicking Breast Phantom Materials, B. Phys. 2017. Juni 04: 31-17. Doi 10.1088/2057-1976/aa7992v.
  19. Asnaeni Ansar1,a, Dahlang Tahir1,b ,Bualkar Abdullah1,c, Nurhasmi1,d , Siti Fatimah1, Jusmawang1,f Physical Characteristics of Soft Tissue Phantom from Silicone Rubber Based Vulcanization System. Materials Science Forum Submitted .2018. Desember 14: Vol. 966, pp 194-199. Doi:10.4028/www. scientific. Net /MSF.966.194.
  20. Y. Wang, B.L. Tai, H. Yu, A.J. Shih. Journal of Medical Devices. Journal. Silicone-Based Tissue-Mimicking Phantom for Needle Insertion Simulation, J.2014. 8(2): 021001 (7 pages). Doi.org/10.1115/1.4026508.
  21. Lestari YM, Anam C, Sutanto H, et al. Biomed Phys Eng Express. Evaluating silicone rubber-lead shield's effectiveness in protecting the breast during thoracic CT. 2023. February3: 9(2). Doi 10.1088/2057-1976/acb551.
  22. Kalra, M.K. Springer-Verlag Berlin Heidelberg.Automatic Exposure Control in Multi-detectorRow Computed Tomography.2017. Januari. DOI:10.1007/978-3-540-68575-3_7.
  23. Gertraud Maskarinec, Lixin Meng, Giske Ursin. International Journal of Epidemiology, Ethnic differences in mammographic densities. 2001. October.Volume 30, Issue 5, Pages 959–965, https://doi.org/10.1093/ije/30.5.959.
  24. Xinyang Sheng, Xiaona Chen and Mark John Lake. Seciendo. Vertical Breast Displacement in Asian Women During Exercise: influence of Bra Type, Size and Different Parts of the Breast. 2022. 12 Oktober: Volume 30. https://doi.org/10.2478/ftee-2022-0016.
  25. Alothmany, N., Molla, N.I., Yusuf, M., Saoudi, A., Mail, N., Alothmany, D., Khafaji, M.A., Natto, H., Tayeb, M., Nadwi, F., Jiman, A., Kinsara, A.A.Radioprotection .Characterization of optically stimulated luminescence for assessment of breast doses in mammography screening. 2016. March 28, vol:51–58. doi.org/10.1051/radiopro/2015032.
  26. Akselrod, M.S., Bøtter-Jensen, L., McKeever, S.W.S., Radiat. Meas. Optically stimulated luminescence and its use in medical dosimetry. 2007. December.Vol: 41, S78–S99. Doi: 10.1016/j.radmeas.2007.01.004
  27. Ketelsen D, Buchgeister M, Fenchel M, et al. Invest Radiol. Automated computed tomography dose-saving algorithm to protect radiosensitive tissues: estimating radiation exposure and image quality considerations. 2012. Feb;47(2):148–152. Doi: 10.1097/RLI.0b13e311504.
  28. Teemu Makel , Mika Kortesniemi a , Touko Kaasalainen. Physica Medica. The impact of vertical off-centering on image noise and breast dose in chest C.T.C.T. with organ-based tube current modulation: A phantom study. Volume 100, August 2022, pages 153-163. doi.org/10.1016/j.ejmp.2022.06.014.

Downloads

Published

2023-12-30

Issue

Section

Research Articles

How to Cite

[1]
Mela Nur Hikmah, Choirul Anam, Heri Sutanto, Oktarina Damayanti, Mohamad Ramdhan, " The Feasibility of Tube Current Modulation (TCM) to Reduce dose of the Surface Breast in Various Breast Sizes, International Journal of Scientific Research in Science and Technology(IJSRST), Online ISSN : 2395-602X, Print ISSN : 2395-6011, Volume 10, Issue 6, pp.402-407, November-December-2023. Available at doi : https://doi.org/10.32628/IJSRST52310682